300-002547-00 rev7 117
ID Description Message
1200 Nav scope EM quality Poor quality navigation field generator data
detected. Navigation accuracy may be
compromised. Press OK to continue.
1201 Poor quality image detected. Navigation may be
compromised.
1202 Poor quality EM data detected. Navigation may
be compromised.
1203 Nav error A navigation error has been detected. Navigation
accuracy may be compromised. Press OK to
continue.
1204 Navigation error detected. Navigation may be
compromised.
1205 Nav initialization error A navigation initialization error has occurred.
Repeat initialization.
1206 Nav initialization error Insufficient EM data for navigation initialization.
Repeat initialization.
1208 EM distortion has been detected. Navigation may
be compromised.
1209 Field generator or patient sensor motion has
been detected.
1210 Patient or FG motion Navigation field generator and/or patient
movement has been detected. Navigation
accuracy may be compromised. Visually inspect
the navigation field generator and/or patient
sensor positioning and initialize navigation at the
main carina.
1211 Nav retraction outside trachea The bronchoscope tip location is outside of the
segmented trachea. Move the bronchoscope tip
closer to the main carina and initialize navigation.
1212 Bronchoscope buckling has been detected.
Check bronchoscope.