118 300-002547-00 rev7
ID Description Message
1213 Scope buckling Bronchoscope buckling has been detected.
Retract and visually inspect the Bronchoscope
System for damage or kinks. Lubricate as
necessary prior to reinsertion. Press OK and
initialize navigation at the main carina.
1214 Nav EM registration error An EM error has been detected. Repeat
navigation initialization. If problem persists,
contact Auris support.
1215 Nav contralateral registration Contralateral registration failed. Repeat
navigation initialization.
1216 Navigation latency detected. Navigation may be
compromised.
1217 Navigation error detected. Navigation may be
compromised.
1300 Bronchoscope System loading
error
The Bronchoscope System failed to load
properly. Unload and reload the device.
1301 Instrument Manager FW
Mismatch
A system error has occurred. All robotic functions
have been disabled. Unload the Bronchoscope
System and reboot the system. If problem
persists, contact Auris support.
1401 EtherCAT Missing Frame(s)
Over Max Limit
A system error has occurred. All robotic functions
have been disabled. To resolve, unload then
reload the Bronchoscope System and click OK to
acknowledge the fault. Then, re-initialize
navigation at the main carina.
1402 EtherCAT Jitter Over Max Limit A system error has occurred. All robotic functions
have been disabled. To resolve, unload then
reload the Bronchoscope System and click OK to
acknowledge the fault. Then, re-initialize
navigation at the main carina.