96 Maintenance
CFPP Process Analyzer CFPP-4.2 03/2016
• Position the decondensation ring (11). Tighten the holder of the decondensation ring
at the frame with two screws (6).
• Push the quick-release couplings of all pipes onto the respective plug nipples with the
corresponding color.
Note: For changing the peltier elements the temperature sensors have not been
dismounted. The screw connections described in the following have not been
loosed!
!
CAUTION!
The tips of the temperature sensors, particularly the sample temperature
sensor (glass tip!) are very fragile and ca
n bend or break if handled
FOR THIS REASON:
- Handle temperature sensors with great care.
- Do not exchange temperature sensors.
• Carefully slide the sample temperature sensor (with glass tip) into the screw
connector.
Fig. 10.19: Tightening screw connections
• Slightly tighten the corresponding union nut while holding the assembly in position
with a wrench.
• Push the connectors for the temperature sensors and for the liquid sensor (5) into the
corresponding sockets with the same color.