Programming the Pump 99
SIGMA Spectrum Infusion System Pump Operating Software Version 8.00
Operator’s Manual For Use With MDL Version 8.0
NOTE: If the standby period is set to infinite, the time value in the display will be replaced
with a dashed line.
3. While the Pump is in standby, press RUN at any time to begin the infusion.
NOTE: Pressing any other key or opening the Pump door cancels standby mode.
4. To start the infusion, ensure that the slide clamp and roller clamp are open and press RUN.
Once RUN is pressed, the infusion will start and the Primary Check Flow screen is displayed.
Check the flow and confirm:
All clamps are open
There are no kinks or collapses in the tubing outside of the Pump
Drops are flowing in the drip chamber. Note, at very low rates, it may take several minutes
to see drops.
Vents are open (if applicable).
If drops are falling in the drip chamber, press the yes soft key.
If no drops are falling, press the no soft key and follow the screen prompts.
To place a Pump in standby after an infusion has started:
1. Press STOP.
2. Press the review/program soft key.
3. Press the hold soft key.
Figure 126. In STANDBY Message.