78 Programming the Pump
Manual 41018v0800 SIGMA Spectrum Infusion System
Revision B Operator’s Manual
NOTE: Values exceeding the hard limits set in the Drug Library will not be accepted by the
Pump.
5. Press RUN to begin delivery at the new flow rate.
Changing the VTBI Without Stopping the Pump
To change VTBI without stopping the Pump:
1. At the Infusion Running screen, press the review/edit VTBI soft key. See Figure 78.
2. At the Review screen, press the edit VTBI soft key. See Figure 79.
3. Enter new value and verify the VTBI is equal to or less than the volume in the container.
See Figure 79.
4. Press OK to confirm VTBI. See Figure 79.
Figure 78. Infusion Running Screen.