EasyManua.ls Logo

baxter 35700BAX2 - Page 83

baxter 35700BAX2
180 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Programming the Pump 77
SIGMA Spectrum Infusion System Pump Operating Software Version 8.00
Operator’s Manual For Use With MDL Version 8.0
If soft limits are exceeded, press the yes soft key to accept the new value or the no soft key to
return to the previous value.
NOTE: Values exceeding the hard limits set in the Drug Library will not be accepted by the
Pump.
5. Observe the change in the mL/hr rate based on the new dose entry.
6. Press RUN to begin delivery at the new dose rate.
To change rate mL/hr without stopping the Pump:
1. At the Infusion Running screen, press the rate change soft key. See Figure 76.
2. Observe the displayed mL/hr soft limits. See Figure 77.
3. Enter a new mL/hr flow rate. See Figure 77.
4. Press OK to confirm new flow rate. See Figure 77.
If soft limits are exceeded, press the yes soft key to accept the new value or the no soft key to
return to the prior value.
Figure 76. Infusion Running Screen.
Figure 77. Rate Change Screen.

Table of Contents

Related product manuals