134 Appendix A: Specifications
Manual 41018v0800 SIGMA Spectrum Infusion System
Revision B Operators Manual
Standards
IEC60601-1, including collateral standards; Third Party Notified Body Testing (Reference
Electromagnetic Compatibility Tables)
IrDA Serial Infrared Physical Layer Link Specification v1.4 (IrPHY)
Wireless – IEEE 802.11b and IEEE 802.11g
Pump Specifications
Specification Description
AC Power AC Power Adaptor, low profile, covers only one outlet, Medical
Grade (EN60601-1-2):
Input (P/N 35727): 115 VAC ±15%, 50 - 60 Hz / 300 mA Max
Output (P/N 35727): 9 VDC/1200 mA, short circuit protected
Cord length 3.0 m (approximately 9.75 ft)
Use only Baxter P/N 35727.
Alarm Volume Variable (three levels: high, medium and low). For more detail, see
Appendix G: “Default Settings” on page 160.
Alarms and Alerts
Air In Line: dual-beam ultrasonic detector alarms for large
bubbles but allows smaller bubbles to pass.
Detects air bubbles >2.5 cm (>1 in) (approximately 140 µL
in Baxter sets)
Detects >1 mL of accumulated air over 15 min., excluding
<10 µL bubbles, at room temperature
Detects >1.5 mL of accumulated air over 15 min., excluding
<10 µL bubbles, at 15.5°C (60°F)
Downstream Occlusion: When the Downstream Occlusion
Automatic Restart is enabled, automatic restart occurs after the
downstream occlusion is cleared. Actuation can be set to:
Low, 41 kPa ±27 kPa (6 ±4 psi)
Medium, 89 kPa ±41 kPa (13 ±6 psi)
High, 131 kPa ±62 kPa (19 ±9 psi)
Very Low Battery: 15 minutes of battery power remaining
Due for inspection: Preventive Maintenance and/or Network
Certification
Anti-Free-Flow System Set based, utilizing IV set slide clamp.