34 Setting Up the Pump
Manual 41018v0800 SIGMA Spectrum Infusion System
Revision B Operators Manual
NOTE: An icon is displayed next to any configuration item in the Options menu (User or
Biomed) where the setting has been assigned in the Drug Library. See Figure 12.
Display Settings
Run Screen Options. Each of the items in this list may be set to On or Off. If enabled, they
are included in the alternating screens that are displayed while the Pump is running.
Audio level indicator – Shows L, M, or H (as selected in Audio Volume).
Rate mL/hr – Shows the mL/hr infusion rate of the current delivery.
Dose rate – Shows the dose rate of the current delivery.
mL VTBI – Shows the remaining Volume To Be Infused, in mL.
Time (hr:min) – Shows the time remaining in the current infusion, in hr:min format.
Display Adjust. The display backlight brightness level may be adjusted from 1 to 10 (10 being
the highest) or set to Off. The backlight consumes approximately 400 mW when set to
maximum brightness; therefore, battery life is maximized when the backlight is set to Off.
To adjust display brightness:
1. Access the User Options menu screen:
From the Care Area selection screen, press the options menu soft key. Then use the
arrow soft keys to select User Options and press OK.
From the Infusion Running screen, press the info settings soft key.
2. Using the arrow soft keys, select Display Settings and press OK. The Display Settings menu
screen appears.
3. Using the arrow soft keys, select Display Adjust and press OK. The Display Adjust screen
appears.
4. Using the arrow soft keys, raise or lower the brightness to the desired level.
5. After selecting the desired brightness level, press OK to save the level and exit the Display
Adjust screen.
6. Press the exit soft key as needed to return to the Care Area selection screen or Infusion
Running screen.
View Information
Pump Information. The Pump Information screen shows the following read-only information:
SW (software) Version
Serial number – The serial number assigned by Baxter for tracking and device history.
Figure 12. MDL Icon.