Setting Up the Pump 35
SIGMA Spectrum Infusion System Pump Operating Software Version 8.00
Operator’s Manual For Use With MDL Version 8.0
Tube type – The name of the IV tube set manufacturer that the device is calibrated for use
with.
Wireless Module
From the Pump Information screen, press the sw info soft key to display the Software Version
screen, which shows the versions of the individual software components that are installed.
SW Version
Sharp
PIC
CPLD
SmartBatt Charger build
Network Module
Library Information. The Library Information screen identifies the name of the Active Drug
Library in Pump memory as well as the Name, Date Modified, Version Number, Format
Indicators, Drugs, Care Areas and Advisories (or modifiers, depending on MDL version).
Show Clinical Advisory. If the current infusion has a clinical advisory associated with it,
selecting this option will display the advisory. If the current infusion does not have a clinical
advisory associated with it, this option is not selectable and appears in grey text in the display.
Infusion Information. The Infusion Information screen identifies the infusion-specific primary
and secondary bag parameters that are not otherwise displayed on the Setup screen. The
information includes the following information for each bag (if programmed): audio level, Near-
Empty Alert status (primary only), pressure setting, KVO rate (primary only) and primary
siphoning alert or secondary complete alert status (secondary only).
History Log. Selecting this option provides access to the event log. Available options for
viewing the log are listed below.
NOTE: Access to this option is provided only when the Pump is not running (delivering). If
the Pump is running, this option is not selectable and appears in grey text in the
display.
View History Log – Select this option to view the entire history log on the Pump screen.
View System Error Log – Select this option to view only the system errors recorded in the
history log.
View Drug Error Log – Select this option to view only the events associated with
programming an infusion; including any drug limits that may have been exceeded during
setup.
Dump History Log – Select this option to send the entire history log out of the Pump using
the IrDA port. Refer to the Logging section of the SIGMA Spectrum Infusion System
Service Manual (P/N 41019v0800).