118 Cleaning and Storage
Manual 41018v0800 SIGMA Spectrum Infusion System
Revision B Operators Manual
PDI
®
Sani-Cloth
®
Bleach Wipes
Super Sani-Cloth® Germicidal Disposable Wipe
Opti-Cide3
®
Surface Disinfectant Wipes
Opti-Cide3
®
Disinfectant and Cleaner Solution
Sklar Disinfectant Solution
DisCide
®
Ultra Disinfecting Towelettes
3M
HB Quat Disinfectant Cleaner Concentrate
MicroKill Plus Germicidal Wipes
Oxivir TB Wipes
Cleaning the Pump
Do not use rigid cleaning instruments or spray solutions directly on the Pump and its accessories.
NOTE: Refer to the facility’s protocol for the frequency of cleaning the Pump.
To clean the Pump:
1. Power off the Pump and unplug the AC Power Adaptor from the power source.
2. Remove the battery module.
NOTE: It is important to remove the battery module to ensure proper cleaning of the battery
terminals, as well as to prevent any power from being applied to Pump circuitry
while liquid cleaning solution is present, which may cause corrosion.
3. If a pole clamp is attached to the Pump, remove the pole clamp by loosening the thumb screw
on the Pump side adaptor (back of the Pump) and slide the Pump off of the pole clamp.
4. Place the Pump in an upright position (keyhole assembly upward).
NOTE: Do not soak any portion of the Pump with liquid cleaners. All cleaning applicators
should be damp, not dripping.
5. Apply the compatible cleaning agent to a lint-free cloth. Use appropriate dilution ratio per the
manufacturer’s instructions.
6. Ring out any excess cleaning solution from the lint-free cloth to ensure it is damp, not dripping.
This prevents fluid from seeping into component areas of the Pump.
NOTE: Disinfectants should remain on the Pump’s surface in an even, but not dripping, film
for the recommended contact time for the compatible cleaning agents.
7. Wipe down the outside of the Pump.