Cleaning and Storage 117
SIGMA Spectrum Infusion System Pump Operating Software Version 8.00
Operator’s Manual For Use With MDL Version 8.0
CLEANING AND STORAGE
The SIGMA Spectrum Infusion System should be cleaned and disinfected for each patient use
according to facility protocol.
CAUTION Cleaning the Pump and Pump Accessories.
Always wear gloves when cleaning Pump and Pump accessories.
Only use Baxter specified compatible cleaning fluids.
Do not allow fluid to seep inside the Pump (especially through the
keyhole, door latches or rear case speaker vent) or severe
damage may occur.
Do not spray solutions directly on the Pump and Pump
accessories.
Do not autoclave or use EtO (ethylene oxide) to sterilize Pumps or
Pump accessories.
Do not apply cleaners directly to battery packs exposed terminals.
Do not immerse any part of the Pump or battery.
Do not use phenol-based cleaners/disinfectants. Phenols degrade
plastics and membrane switches.
Do not use abrasive cleaners.
Do not use rigid cleaning instruments.
Always use a lint-free, foam tipped swab to clean the tube channel.
Always dispose of all cleaning materials per federal, state and
local regulations for biohazard waste disposal.
Compatible Cleaners
For a complete list of compatible cleaning materials refer to DOC 11318 on the Baxter website
(www.sigmapumps.com).
■ 10% Solution of Clorox
®
Regular-Bleach and 90% Water
■ Clorox Commercial Solutions
®
Clorox
®
Germicidal Wipes
■ Up to 90% Isopropyl Alcohol (IPA)
■ Dispatch
®
Hospital Cleaner Disinfectant Towels With Bleach
■ Coverage
®
Plus Germicidal Surface Wipes
■ Coverage Spray TB Plus Ready to use Disinfectant and Decontaminant
■ CaviCide
®
Disinfectant/Decontaminant Cleaner
■ CaviWipes
®
Disinfectant/Decontaminant Cleaner
■ EnVerros Sanimaster 4
®
■
A-456 II Disinfectant Cleaner
■ PDI
®
Sani-Cloth
®
AF Germicidal Disposable Wipe