Programming the Pump 93
SIGMA Spectrum Infusion System Pump Operating Software Version 8.00
Operator’s Manual For Use With MDL Version 8.0
The Infusion Running screen will show the amount of volume remaining and an indicator that
the line is being flushed. See Figure 111.
The Line Flush will continue until the entered flush volume has been delivered or the air in line
detector senses air. See Figure 112.
5. When no additional flushes are available (maximum of 3), a Select Infusion screen displays to
return to the Secondary Setup or Review Primary infusion. See Figure 113.
Changing the Rate of an Amount/Time Infusion
NOTE: During an Amount/Time Infusion, a dialog box is displayed if the limits defined in the
Drug Library for the selected drug are exceeded.
Values exceeding the Soft Limits can be accepted by pressing the yes soft key or declined by
pressing the no soft key displayed in the dialog box.
Hard limits cannot be exceeded. Re-enter values within hard limits.
Figure 111. Secondary Line Flush Status
Screen. See important note on page 91.
Figure 112. Secondary Line Flush Complete
Screen.
Figure 113. Select Infusion Screen.