Appendix G: Default Settings 161
SIGMA Spectrum Infusion System Pump Operating Software Version 8.00
Operator’s Manual For Use With MDL Version 8.0
BASIC Configurations
A BASIC infusion provides non-DERS based infusion programming.
DERS limits do not exist when using BASIC mode. BASIC mode allows the user to manually
specify an mL/hr rate, dose mode, dose rate and volume to be infused, among other parameters. It
does possess the following safety features: Check Flow at Run, Secondary Error Prevention and
Single Step Rate Change of 101%.
CAUTION BASIC Programming Use.
BASIC programming should only be used when the desired drug or
concentration is not available in the facility’s Drug Library.
Table 3. BASIC Fixed Settings.
BASIC Configuration Settings Fixed Settings
Allow Bolus Yes
Allow Loading Dose No
Bolus Amount Units mL
Bolus Time Units Minutes and Seconds (00:00)
Completion Alarm Not Available
Delayed Run Not Available
KVO Rate 1 mL/hr
Secondary Callback Never
Single Step Rate Change 101%
Table 4. BASIC Settings Configurable on the Pump.
BASIC Configuration Settings Default Settings
Audio Alarm High, Medium, Low
Bag Near Empty Alarm On, Off
Delivery Bag Primary Bag, Secondary Bag
Dose Mode Clinician can select