Cleaning and Storage 121
SIGMA Spectrum Infusion System Pump Operating Software Version 8.00
Operator’s Manual For Use With MDL Version 8.0
Cleaning the Pole Clamp
To clean the pole clamp:
1. Power off the Pump and unplug the AC Power Adaptor from the power source.
2. Loosen the thumb screw on the back of the Pump and slide the Pump off of the pole clamp.
3. If the pole clamp is still attached to an IV pole, turn the mounting knob counter clockwise to
remove.
4. Apply cleaning agent to a dampened cloth per the manufacturer’s instructions, using
appropriate dilution ratio.
5. Wipe the entire pole clamp including the screw mechanism.
6. Use a soft brush to clear debris from the screw mechanism.
7. Dry the pole clamp thoroughly before replacing it onto the Pump.
8. Dispose of all cleaning materials as required per facility protocol/biohazard policy.