74 Programming the Pump
Manual 41018v0800 SIGMA Spectrum Infusion System
Revision B Operators Manual
To end the multi-step program:
» At anytime during the infusion or, at the end while infusion complete displays, press STOP and
press the clear program soft key to clear the entire program.
To retain the existing program in memory and restart at the same step, or after infusion
complete (returns to Step 1):
1. Press STOP.
2. Press ON/OFF to power off the Pump.
When the Pump is powered back on, the same program will be retained in memory until the
clear program soft key is pressed.
To clear the entire multi-step program:
»Press the clear program soft key and answer yes to the confirmation screen.
To clear the last step in the multi-step program:
1. Ensure that the Pump is stopped.
2. Press the review or program pri/sec soft key.
3. Press the arrow soft keys to scroll to the desired step in the step bar.
NOTE: Press the clr step soft key. The last step of the program is the only step that can be
cleared.
Cyclic TPN
Cyclic TPN delivers TPN (Total Parenteral Nutrition) over a prescribed period of time in an
automatic ramp-up infusion rate to a main rate before tapering down. This process increases the
delivery rate at the start of an infusion and decreases the delivery rate at the end of the infusion.
The Pump calculates the ramp-up and taper-down rates once the duration and total volume of the
infusion are entered. The infusion schedule is calculated containing the rates and volumes for:
10 ramp-up steps consisting of 10% of the total infusion time.
The main delivery rate accounts for 80% of the infusion time.
10 taper-down steps for the remaining 10% of the total infusion time.
Cyclic TPN is only available for Continuous Infusions.
To program Cyclic TPN: