Programming the Pump 79
SIGMA Spectrum Infusion System Pump Operating Software Version 8.00
Operator’s Manual For Use With MDL Version 8.0
5. Press RUN to begin delivery of the new VTBI.
Delayed Run
If Delayed Run is enabled (per drug in the Drug Library), the start of any programmed infusion may
be delayed up to 12 hours.
To set up a delayed run:
1. On the Infusion Setup screen, enter any value between 1 minute and 12 hours (00:01 to
12:00, hr:min) for the delay run parameter and press OK.
2. After the infusion programming is complete and the set is loaded, press RUN to begin the
infusion delay timer. The delay running screen is displayed with the remaining delay time
shown in a flashing format.
3. When the delay time period expires, the Pump begins delivery of the programmed infusion.
Primary Check Flow will display.
4. An inactivity alarm will occur after 2 minutes and Primary Check Flow confirmation will be
required.
To stop or cancel the delay timer:
While the delay timer is running, it can be stopped or cancelled.
Figure 79. VTBI Review Screens.