Setting Up the Pump 33
SIGMA Spectrum Infusion System Pump Operating Software Version 8.00
Operator’s Manual For Use With MDL Version 8.0
To access the User Options menu:
From the Care Area selection screen:
1. Press the options menu soft key.
2. Select User Options from the menu and press OK.
From the RUN or STOPPED screen:
» Press the info settings soft key.
The User Options menu has three categories, which are described in the sections that follow:
Alarm Settings
Display Settings
View Information
Alarm Settings
Audio Volume. The audio volume of the Pump has three levels: LOW, MEDIUM and HIGH. To
adjust the Audio Volume, use the arrow soft key to select the desired volume level and press
OK.
Audio Volume may be preset in the Master Drug Library. The default setting is “Use Pump
Setting”, which is adjustable at the Pump.
Audio Tone. The audio tone has two settings: short and long. To adjust the audio tone, use
the arrow soft key to select the desired tone and press OK.
Standby (hr:min). Set the length of time to keep the Pump in Standby (or Hold) after setup of
the infusion has been completed. Settings are from 00:01 to 99:59 (hr:min). Setting the value
to 00:00, or Infinite, results in an infinite Hold period.
Near Empty alert. When enabled, a Near Empty alert is displayed when 30 minutes or less of
infusion remains. If the initially-programmed infusion is less than 30 minutes, the alert is not
generated, regardless of the setting.
Near Empty alert has three options in the Drug Library: On, Off and “Use Pump Setting”. The
default setting is “Use Pump Setting”, which is adjustable at the Pump.
DS Pressure Limit (Downstream Occlusion Pressure Limit). The DS Pressure limit has
three levels:
Low, 41 kPa ±27 kPa (6 ±4 psi)
Medium, 89 kPa ±41 kPa (13 ±6 psi)
High, 131 kPa ±62 kPa (19 ±9 psi)
To adjust the DS Pressure Limit, use the arrow soft key to select the desired level and press
OK.
The default Pressure Limit may be preset for a Care Area in the Drug Library. The default
setting is “Use Pump Setting”, which is adjustable at the Pump.