122 Cleaning and Storage
Manual 41018v0800 SIGMA Spectrum Infusion System
Revision B Operators Manual
Pump Handling, Transport and Storage
Do not handle, transport or store Pumps in ways that might result in physical damage.
Avoid handling, transporting and storing Pumps in a manner in which the pole clamp of
another Pump, or other heavy or sharp objects could impact the keypad.
WARNING Pump Storage.
Remove the Battery Module from the Pump when storing the Pump
for extended periods.
WARNING Ensure Secure Mounting of Pump During Use and Transport.
During use and transport, securely mount Pumps to IV pole by
centering the pole in the clamp and turning the mounting knob
clockwise. To maintain IV pole stability, never exceed 210 cm (83 in)
from floor to IV pole top and limit bag volume at this extended height
to <1 liter (1000 cc).
Proper Pump Handling, Transport and Storage.
Improper Pump Handling, Transport and Storage.
Figure 135. Pump Handling, Transport and Storage.