Programming the Pump 101
SIGMA Spectrum Infusion System Pump Operating Software Version 8.00
Operator’s Manual For Use With MDL Version 8.0
NOTE: 429 is the default keypad lock code. Refer to the Library Configurations section in
the SIGMA Spectrum Infusion System Master Drug Library User Manual
(P/N 41020v080) to change the keypad lock code. The keypad lock code is set
hospital-wide in the Drug Library and may be 1 to 4 digits long (1-9 only; zero cannot
be used).
To unlock the keypad:
» Re-enter the keypad lock code 429. The Keypad Lock icon disappears.
NOTE: If the keypad is unlocked while reviewing setup and the Pump is not stopped, the
keypad will automatically relock upon returning to the Infusion Running screen.
NOTE: Pressing the STOP key while the keypad is locked will not stop the Pump. To stop
the Pump, re-enter the keypad lock code and press STOP.
To view infusion setup while the keypad is locked:
» Press the review soft key or the review/edit VTBI soft key. No values can be changed and,
therefore, navigation from value to value is not allowed while the keypad is locked.
Certain alarm conditions can be silenced and cleared while the keypad is locked.
For more information about the keypad lock feature, refer to the Care Area Configurations section
in the SIGMA Spectrum Infusion System Master Drug Library User Manual (P/N 41020v080).