Alarms 111
SIGMA Spectrum Infusion System Pump Operating Software Version 8.00
Operator’s Manual For Use With MDL Version 8.0
SYSTEM ERROR Displays when the Pump detects an internal fault.
NOTE: Record the programming parameters before the Pump is powered off.
Follow the directions displayed on the screen to:
■ Power off the Pump and then power on the Pump.
Should an audio alarm occur with no direction on the screen:
1. Disconnect the battery from the Pump and unplug the AC Power Adaptor from
the AC outlet.
2. Wait approximately 5 or more seconds, reconnect the battery and plug the AC
Power Adaptor into an AC power outlet.
3. Press ON.
4. If neither procedure clears the fault, return the Pump for service.
CAUTION Unrecoverable System Error.
If unable to clear a fault condition during a
system error occurrence, discontinue using
the Pump. Refer to qualified service
personnel at your facility or return the Pump
to Baxter for service.
Alarms Action