Introduction and Safety 3
SIGMA Spectrum Infusion System Pump Operating Software Version 8.00
Operator’s Manual For Use With MDL Version 8.0
WARNING Battery Removal.
Do not detach the battery during patient therapy.
Never touch the patient and the Pump at the same time with the
battery removed and the Pump connected to the power outlet.
WARNING ESD Sensitivity.
Do not touch the battery pin set when the Battery Module is removed.
WARNING Adjacent or Stacked Use.
The Pump should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the Pump should
be observed to verify normal operation in the configuration in which it
will be used.
NOTE: The SIGMA Spectrum Infusion System has been tested to
operate normally when used stacked or adjacent to other
SIGMA Spectrum Infusion Systems.
WARNING Pump Storage.
Remove the Battery Module from the Pump when storing the Pump
for extended periods.
WARNING Proper Disposal Required.
To dispose of this device or the associated administration sets,
adhere to local, state, federal and/or other governing regulations.
WARNING Magnetic Fields.
The SIGMA Spectrum Infusion System is not designed to be MRI-
compatible nor is it intended to be used in this manner. Strong
magnetic fields (those beyond the level tested) may cause the device
to operate improperly.
Do not expose the Pump to strong magnetic fields such as is
common with MRI equipment or in close proximity 60.9 cm (2 ft) of a
cathode ray tube (CRT) monitor. Doing so may cause injury to the
patient and/or damage to the equipment.
WARNING Linear Accelerator Radiation.
The SIGMA Spectrum Infusion System is not designed to be exposed
to linear accelerator radiation nor is it intended to be used in this
manner. Exposure to radiation of this type may cause the device to
operate improperly.
Do not expose the SIGMA Spectrum Infusion System to linear
accelerator radiation. Doing so may cause injury to the patient and/or
damage to the equipment.