3 Disassembly / Assembly
Infusomat® compact
plus
P 1.0 3 - 13
EN
For internal use, only
4. Carefully lift the upper part of the unit (Fig. 3 - 14 / Item 1),
swing it over to the left and set it down.
5. Remove the fixing plate (Fig. 3 - 14 / Item 3) from the lower
part of the unit (Fig. 3 - 14 / Item 4).
6. Disconnect the connector of the ESD wire (Fig. 3 - 15 / Item 1)
on the processor board.
7. Disconnect the connector of the power supply (Fig. 3 - 15 /
Item 2) on the processor board.
8. Disconnect the connector of the stepper motor (Fig. 3 - 15 /
Item 3) on the processor board.
9. Completely unlock the connector for the connecting cable of
the sensor board and pull out the ribbon cable
(Fig. 3 - 15 /
Item 4).
Fig. 3 - 14
1 Upper part of the unit
2 Gasket (all-round)
3Fixing plate
4 Lower part of the unit
1
3
4
2
Fig. 3 - 15
1 Connector for ESD wire
2 Connector for power supply
3 Connector for stepper motor
4 Connector for sensor board
4
3
21
NOTICE
Check the gasket (Fig. 3 - 14 / Item 2) for integrity. In case of
any visible damage replace the housing, upper part.