4 Check after Repair
4 - 12 Infusomat® compact
plus
P 1.0
EN
For internal use, only
T8 Second clamp
1. Manually open the SHK
2. Remove the infusion line from the SHK.
Still leave the infusion line in the second clamp Fig. 4 - 8!
3. Read off the "P
2ndClamp
" value on the pressure gauge, compare
with the specifications in the checklist, and document.
4. Dispose of the infusion line (the injection port may be used
again).
T9 Delivery accuracy (simple method)
Requirements:
■ Test setup according to Fig. 4 - 2, with the liquid level
approx. 50 cm above the pump
■ Ambient temperature according to the checklist
■ Target volume: 25 ml
Measurement:
1. Set a volume of 30 ml (VTBI) to reduce possible spilling.
2. Insert the Intrafix Primeline of the test setup (Fig. 4 - 2) in the
device.
Do not stretch the line when inserting.
3. Close three-way cocks to the syringe and the measuring device,
please see
Fig. 4 - 9.
4. Vent test setup. Bottom part of drop chamber must be 2/3
filled.
5. Set delivery rate according to the checklist.
6. Let device deliver for ≥5 min into a vessel for waste fluid (not
into the measuring container). Then, stop the pump.
7. Insert cannula in empty graduated cylinder and at the same
time start the stop watch and the Infusomat® compact
plus
P.
8. Stop the stop watch when the target volume is reached.
NOTICE
In case the pressure does not reduce as expected, check if the
roller clamp has been closed, accidentally.