70
Error
Code
Error Message Error Resolution
888
Unable save generated JPEG image (unable
validate if the data was actualy saved to
storage)
1. Select Reset Error [F5].
2. Select the X icon in the top
right corner of the ReadA
application title bar to close
the application.
3.
Select the shortcut on the
desktop to re-launch the
ReadA application and sign
in.
4. Select Start [F2] to restart
the module.
900 Fast integrity failed
Contact BD.
970
General device error: Camera and or lights are
not available
974 Requested dish ... is being processed now!
1. Select Reset Error [F5].
2. Select the X icon in the top
right corner of the ReadA
application title bar to close
the application.
3.
Select the shortcut on the
desktop to re-launch the
ReadA application and sign
in.
4. Select Start [F2] to restart
the module.
980 Unknown imaging error Contact BD.
986 Exception while storing SHQI data
1. Select Reset Error [F5].
2. Select the X icon in the top
right corner of the ReadA
application title bar to close
the application.
3.
Select the shortcut on the
desktop to re-launch the
ReadA application and sign
in.
4. Select Start [F2] to restart
the module.
991 Dish was not removed in time
999 General network communication error
BD Kiestra™ ReadA User's Manual