EasyManua.ls Logo

BD Kiestra ReadA - BD Kiestra Optis Error Messages

BD Kiestra ReadA
90 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
68
Error
Code
Error Message Possible Cause Error Resolution
E230
Outfeed: conveyor
motor current limit
overload
Possible obstruction.
1. Remove obstruction.
2. Select Reset Error [F5].
3. Select Start [F2] to restart the
module.
6.4 BDKiestra™ Optis™ error messages
Error
Code
Error Message Error Resolution
200
For unknown reason failed to process client
request
1. Select Reset Error [F5].
2. Select the X icon in the top
right corner of the ReadA
application title bar to close
the application.
3.
Select the shortcut on the
desktop to re-launch the
ReadA application and sign
in.
4. Select Start [F2] to restart
the module.
800 Unexpected error
801 Camera device is not connected
Contact BD.
802 Lights controls are not connected
804 No connection to database
805
Missing required plugin for communicating with
the database
806 Missing camera definition & settings files
807 Camera requires calibration
Recalibrate using the ReadA
application. Refer to
5.4.4.2Calibrating the
BDKiestra™ Optis™ camera.
BD Kiestra™ ReadA User's Manual

Table of Contents

Related product manuals