6.3 Error messages
Error codes, error messages, possible causes, and error resolutions are listed in the following
table:
Error
Code
Error Message Possible Cause Error Resolution
E002
There is not enough
air pressure
The pressure is below
the minimum required
pressure.
Check the air supply.
E003
Watchdog error
occurred
The PLC and PC no
longer communicate.
Contact BD.
E004
Possible broken fuse
detected
Possible broken fuse. Contact BD.
E008 HEPA filter warning
HEPA filters in the
ReadA must be
replaced within 2
months.
Contact the company responsible
for replacement of HEPA filters in
your lab. If the HEPA filter is not
replaced in time, the module will
shut down.
Also contact BD.
E009 HEPA filter error
HEPA filter can no
longer be used.
The module has shut down.
Contact the company responsible
for replacement of HEPA filters in
your lab.
Also contact BD.
E020
Container store:
cylinder not in
position
Possible obstruction.
1. Remove obstruction.
2. Select Reset Error [F5].
3. Select Start [F2] to restart the
module.
BD Kiestra™ ReadA User's Manual