SAMPLE ANALYSIS
PATIENT/SAMPLE INFORMATION
10. If a flagging set is not selected for a sample and either a Worklist is not used or a Worklist
is used but the patient’s age (or date of birth) and sex were not entered on it, the
instrument uses the Default Flagging Set as shown above.
a. The procedure for changing this setting,
Setting a Default Flagging Set, is under
Heading A.10 FLAGGING SETS SETUP, in the Online Help System or the Instructions for
Use manual.
b. To view the default setting, from the Main menu click tt .
Optional Patient/Sample Information (Entered via a Worklist)
1. If a Worklist is used, additional information can be associated with a sample.
a. A Worklist contains sample and patient identification information for specimens
that are pending analysis.
b. Before the sample is analyzed, you can enter or edit patient demographics on the
Worklist.
FLAGGING SET HIERARCHY
No
Current
Flagging set based
on Age or Gender
?
Yes
Yes
No
1
5
YesNo
Age
14yr
?
YesNo
Gender
available
?
Start
7367518B
Select child
range
WomanMan
Flagging set
remains as
selected
Current
Flagging set=
Default Flagging
Set?

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Beckman Coulter COULTER ACT 5diff and is the answer not in the manual?

Beckman Coulter COULTER ACT 5diff Specifications

General IconGeneral
BrandBeckman Coulter
ModelCOULTER ACT 5diff
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals