3-6
SETUP OPTIONS
QUALITY ASSURANCE SETUP OPTIONS
b. XB/XM Options
1) Allow you to enable or disable the XB or XM quality control options.
2)
OFF is the default setting.
3) Information about the XB/XM method of monitoring instrument performance
is available in the Topic,
XB/XM, in Chapter 4.
4) To change the selection see the procedure,
XB/XM Options (Enabling/Disabling), in
Appendix
A.5.
c. Minimum Runs for Automatic Calibration
1) Allows you to enter the minimum number of runs the laboratory will accept.
2) The number 5 is the default.
3) Information about calibration is available in the Topic,
Calibration, in Chapter 5.
4) To change the setting see the procedure,
Minimum Runs Required for
Auto-Calibration (Defining)
, in Appendix A.5.
d. IQAP ID
1) Allows laboratories participating in the IQAP program to enter or change the
IQAP ID received with their enrollment confirmation information.
2) Information about the IQAP method of monitoring instrument performance is
available in under Heading 4.5, INTERLABORATORY QUALITY ASSURANCE
(IQAP) in this Training guide.
3) To enter or change the setting see the procedure,
IQAP ID (Entering/Editing), in
Appendix
A.5.
e. CV Limits (Calibration, QC, Reproducibility)
1) Calibration results are compared to the CV limits listed for calibration.
2) To ensure the instrument is performing to specifications, it is strongly
recommended that you leave the Calibration CV limits set to the default values.
3) The values in the cell control files are compared to the CV limits listed for QC.
4) To change the CV limits for controls, see the procedure,
CV Limits for Calibration,
QC, and Reproducibility
, in Appendix A.5.
5) Reproducibility results are compared to the CV limits listed for Reproducibility.
6) To ensure the instrument is performing to specifications, it is strongly
recommended that you leave the Reproducibility CV limits set to the
default values.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Beckman Coulter COULTER ACT 5diff and is the answer not in the manual?

Beckman Coulter COULTER ACT 5diff Specifications

General IconGeneral
BrandBeckman Coulter
ModelCOULTER ACT 5diff
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals