PN 177196BB
3-17
SETUP OPTIONS
FLAGGING SETS SETUP OPTIONS
3
a. Click the two tabs alternately, noting that only the data on the right of the screen
changes, switching from patient and action limits to flag sensitivity and
thresholds settings.
b. The data displayed on the right is always for the flagging set selected on the left.
c. The options in the Flag Sensitivity and Thresholds box are for Service use only.
d. The options in the Flags and Messages box allow you to check, and edit if desired,
the patient limits and action limits for the flagging set selected.
e. In the Flagging Sets box on the left side of the screen, click on one of the Child
flagging sets.
f. Note that now a third tab,
Age Ranges, is displayed.
4. Click on the
Age Ranges tab.
a. The Age Ranges box allows you to edit the patient age defined for
Child 1, Child 2,
Child 3, Child 4, and Child 5.
b. You cannot type an age into the age field.
1) You use the slider to change the age.
2) You press è or é to adjust the age by one day.
c. You cannot make the next flagging set younger than the former.
For example,
Child 2 cannot be younger than Child 1.
5. In the Flagging Sets box, have the trainee select each flagging set and check the patient
limits, the action limits, and the defined age where it applies, for that flagging set.
a. If any of the flagging sets need to be changed, or more flagging sets need to be
added, help the trainee find and use the appropriate procedures under
A.10,
FLAGGING SETS SETUP
, in the Online Help System or the Instructions for Use manual,
to make the changes.
b. If the flagging sets meet the laboratory’s current needs, just make sure the trainee
knows where to find the procedures for setting up and editing the flagging sets.
Help them find
A.10, FLAGGING SETS SETUP, in the Online Help System or the
Instructions for Use manual.
IMPORTANT Do not make any adjustments to the flag sensitivities or the thresholds without first
consulting your Beckman Coulter Representative. Otherwise, your system may not perform to
specifications. Any change to thresholds or sensitivity affect overall system performance.
IMPORTANT Due to the system’s calculation methods for determining the age from the date of birth, the
precision of the age calculation is limited to ±1 day. When the age is close to the limit of a flagging range,
the adjoining flagging range may be selected.