8. Alarms and Troubleshooting Guide
8-107
3. FLUID OUT
If the blood or replacement fluid runs out while infusing or if the tubing in the fluid out
detector is pulled or kinked, the pump and heating will stop. The Fluid Out message will
display and an audible alarm will sound (figure 4). The valve will go to the recirculate
position.
Figure 4. Fluid Out Alarm
Figure 5. Fluid Out alarm after pressing REPRIME
Press MUTE to silence the alarm
Clamp off the empty bags and bag spikes and replace with new fluid bags.
Pierce new bags and unclamp bag spikes. Open the door to let fluid into the
reservoir before press REPRIME.
The system will automatically stop and return to the Infuse screen after Reprime.
Detection of air during the process will restart the 100 ml cycle.