3. Operation & System Set-up
3-32
PROTECTION FROM ACCIDENTAL POWER OFF
Figure 8. Circuit breaker was turned to STANDBY while pumping.
If the circuit breaker was turned to the STANDBY position while the system is pumping,
the system will stop pumping, alarm and display, Figure 8. This message is to protect
the system from being accidentally powered down during a procedure. To power down
the system, press POWER OFF. To continue with the procedure, turn the circuit breaker
back to the ON position and resume operation.
Clamp off the patient line and bag spikes at the end of the procedure. Turn the system
to STANDBY, using the circuit breaker, first then remove the disposable set from the
system. Make certain the patient line is clamped shut when the door of the system is
open while the system is powered off to prevent uncontrolled fluid flow. Follow the
cleaning procedures outlined in Chapter V to clean and disinfect the system. Practice
standard hospital policy when handling and disposing the bio-hazardous materials.
CAUTION:
With fluid in the disposable set and
the system not powered on, keep
the patient line clamp closed when opening the door to prevent
uncontrolled fluid flow.