3. Operation & System Set-up
3-12 12
STEP
-
BY
-
STEP SUMMARY OF OPERATING PROCEDURES
The following is a step-by-step summary of the major steps for proper operating the
FMS2000; the remainder of the chapter explains each step in detail.
SET-UP
o Inspect the system
Power cord: use only supplied power cord
Reservoir Support
Disposable Set
3.0 L Reservoir, if needed
Large Reservoir Holder, if needed
o IV Pole Mounting
IV Pole: 5-wheel (max diameter 1 ¼”, min 1”)
IV Pole Mounting Support Assembly
o Installing Disposable Set and 3.0 L Reservoir (if needed)
Install reservoir support or large reservoir holder
Install disposable set and 3.0 L Reservoir, if needed
Close the door
Completely close all bag clamps
o Access Power
Plug power cord into the unit
Press circuit breaker to ON. Wait for PRIME screen to appear
o Installing Fluid Bag
Hang fluid bag on IV pole
Remove bag spikes, insert into fluid bag
Open bag clamps
PRIMING
o Priming Main System with solution compatible with blood
Press PRIME - recirculate fluid at 500 ml/min, stop
automatically after 100 ml (countdown is displayed on screen)
o Priming the Patient line
Press PT. LINE PRIME
Press once, prime at 50 ml/min. Press and hold, prime at 200
ml/min
Press STOP after no air in patient line
INITIATING
INFUSION
o Connect to the Patient
Select an appropriate cannula size for desired flow rate
Using aseptic technique, make patient connection without
entrapping air
o Infuse
Press INFUSE to start infusing at 10 ml/min.
Adjust flow rate, as needed
MAINTAIN
INFUSION
o Maintain Infusion
Routinely check patient and system parameters, on screen
Respond to and correct system alarms
BATTERY
OPERATION
o Battery Operation
Press RECIRC key to preheat fluid in the reservoir
Remove power cord from receptacle
END OF
PROCEDURE
o End of Procedure
Clamp off patient line and bag spikes
Turn circuit breaker to STANDBY
Remove disposable set
Dispose of disposable set using hospital policy when handling
and disposing the bio-hazardous materials