8. Alarms and Troubleshooting Guide
8-122
PROBLEM POSSIBLE CAUSE SYSTEM RESPONSE OPERATOR ACTION
Unable to calibrate
temperature probes
Temp probe malfunction
Incorrect fluid temperature
used for calibration.
NA Check the temperature of fluid and make certain it is correct.
If problem persists, service machine – replace temperature probe.
Unable to turn the
system power to OFF
Q11 on the Daughter board
shorted
Unplug from the wall
outlet, not turning off.
Run in the battery mode
until shut down.
Service machine if problem persists – replace the Daughter board.