EasyManua.ls Logo

Belmont FMS2000 - Page 35

Belmont FMS2000
134 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3. Operation & System Set-up
3-27
Figure 6. Patient Line Prime
CONTROLS
PATIENT LINE
PRIME
Primes the patient line with warm fluid. Press the key once to
prime at 50 ml/min or press and hold the key to prime at 200
ml/min.
WARNING! The operator must make certain that the entire fluid
path is completely primed and free of air before administering
fluid to the patient. Failure to do so may result in the introduction
of air into the patient.
STOP
Stops pumping and heating. The system is in standby, ready to
continue on request.
INFUSE
After properly priming the system, press INFUSE to go to the
main operating screen and start infusion at 10 ml/min.
WHEN PT. LINE PRIMED PRESS
STOP AND THEN INFUSE
STOP
PT LINE
PRIME
INFUSE

Table of Contents

Other manuals for Belmont FMS2000

Related product manuals