Tray mat (Option)
Doctor control panel
Doctor Instruments holder
CARE AND MAINTENANCE
Tray mat (Optional)
• The tray mat may be detached and sterilized in an autoclave
(for 20 minutes at 121°C or 12 minutes at 132°C). Remove
residues and clean carefully with running water before
sterilization.
• Be sure to put into a sterilization pouch when sterilizing.
• The tray mat is made of resin and may become deteriorated
if it is sterilized in an autoclave many times. The allowable
sterilization frequency in an autoclave is 250 times.
• After autoclave sterilization, the tray may is subject to
discoloration, which does not have a negative effect on
performance.
Do NOT spray directly
Doctor control panel
•
In case the surface of the doctor control panel is
wiped with disinfectant agent or the like, fully wipe
away the chemical. Control panel malfunction
may arise if the chemical penetrates to the back face
of the doctor control panel.
•
Do not spray liquids directly onto unit surfaces,
holder and operation panel. In order to prevent
damage to electrical components and system.
• Use FD333 or FD366 made by Durr or ethanol as sprayed
to a soft cloth for cleaning and disinfection. Wipe off with
dry soft cloth to dried up for the doctor control panel.
54
2
Doctor Unit Section
CAUTION