Syringe body
Nut
Nozzle
Cleaning Belmont 77 Syringe Nozzle
Cleaning brush
or Tooth brush
Disassembly
CAUTION
Cleaning must be done every after use to patients.
Cleaning must be done within 1 hour after use. Replace with a new syringe nozzle if contamination and
solid material attached to syringe nozzle can not be removed.
Cleaning of Belmont 77 syringe
[Disassembly]
Remove the nozzle from syringe by turning it in direction A.
[Cleaning by hand]
1.Wipe off the surface contamination by a cloth while rinsing
the surface by running clean warm water at 40±5 .
Scrub the tip and joint part of nozzle by a cleaning brush or
by a tooth brush with running clean warm water at 40±5 .
2. Check whether contamination is removed or not after
cleaning. Continue the cleaning if contamination is remained.
3. Immersed with an alkaline disinfection or detergent for
5 minutes. (We recommend to use ID212 made by DURR)
4. Rinse thoroughly by distilled water at ordinary
temperature or by clean water for more than 1 minute.
A
B
w
A
Syringe body
Nut
Nozzle
Assembly
Sterilization pouch
Sterilization of Belmont 77 syringe
The nozzle can be sterilized with autoclave. Sterilization
with autoclave is permitted up to 250 times.
A. Insert the handpiece in a sterilization pouch and seal it.
B. Autoclave for 3 min. at 134 and dry for 15 min.
Apply Vaseline thinly and evenly to the two O-rings after
sterilization. Tighten the nut ¿UPO\ in direction B to undo the
nozzle.
Storage
After sterilizing the nozzle, keep it in the sterilization pouch
and store in a dark and cool place.
CAUTION
• Sterilization must be done every after use to patients.
• Sterilization by class B cycles.
• Sterilization temperature is 135 or less.
• Dry naturally if the temperature for drying process is to exceed 135.
• If damage occurs to the sterilization pouch, discard and sterilize again
using a new pouch.