V6 Vital Signs Monitor Users Manual
3
A.3.3 Battery
Type
Rechargeable lithium ion battery
Size
105m78mm×20mm
Weight
<360 g
Quantity
1
Rated voltage
11.1 VDC
Capability
4000 mAh
Operating time
8 hours
Using a new and fully charged battery at 25 ambient
temperature, connecting SpO
2
sensor and NIBP work on
AUTO mode for 15 minutes interval.
Charge time
6h to 100%Standby
Turn off delay
5 min -15 min after the low battery alarm first occurs.
Indicator of battery
capability
With
A.3.4 Mainframe LED
Physiological alarm
indicating lamp
1 (Yellow/Red)
Technical alarm indicating
lamp
1(Blue)
Power indicating lamp
1(Green/Orange)
GreenWhen powered with AC, it lights green while turn on
and off the monitor.
Orange When powered with battery, it lights orange only
while turn on the monitor.
Battery charging indicating
lamp
1 (Orange)
A.3.5 Audio indicating
Speaker
Gives audible alarm, QRS tone
Supports Pitch Tone and multi-level volume;
Alarm tones meet the requirement of IEC 60601-1-8.
Alarm pressure
45 dB to 85 dB. Testing place is 1 meter from the tone.