SECTION 3 OPERATING THE BIS VISTA MONITORING SYSTEM
______________________________________________________________________
3-2
WARNINGS!
THE CONDUCTIVE PARTS OF ELECTRODES OR SENSOR
AND CONNECTORS SHOULD NOT CONTACT OTHER
CONDUCTIVE PARTS, INCLUDING EARTH.
TO REDUCE THE HAZARD OF BURNS DURING USE OF
HIGH-FREQUENCY SURGICAL EQUIPMENT, THE SENSOR
OR ELECTRODES SHOULD NOT BE LOCATED BETWEEN
THE SURGICAL SITE AND THE ELECTRO-SURGICAL UNIT
RETURN ELECTRODE.
THE SENSOR MUST NOT BE LOCATED BETWEEN
DEFIBRILLATOR PADS WHEN A DEFIBRILLATOR IS USED
ON A PATIENT CONNECTED TO THE BIS VISTA SYSTEM.
TO MINIMIZE THE RISK OF PATIENT STRANGULATION,
THE PATIENT INTERFACE CABLE (PIC) MUST BE
CAREFULLY PLACED AND SECURED.
3. Secure the BISx
Using the attachment clip, secure the BISx to a convenient location near the patient's head.