9
Selection
Subselection Display Explanation
or
33 1
Digital filter
0
No-motion environment, brief weighing time
Setting changes with
each keystroke.
1
No-motion environment, medium weighing time
to
9
Environment in motion, long weighing time
or
34 15
Time to establish no-motion of weight
Time = value x 30 ms
Minimum input = 8
or
36 1
taring
Only with key on the device.
1
100% released
0
0% released
or
38
Portion control
1
Function active
0
Function not active
or
39
Minimal weight change for another recording
Requirement for this is: Step 5 is active (= 1).
0001
0.001 kg
Setting changes with
each keystroke.
to
0098
0.098 kg
9.2 Scale parameter
Slide switch to calibrated. Scale data cannot be
changed.
Change scale parameters: You must first remove
the verification seal to invalidate the verification
of the scale.