Service menu Operating instructions eS10
86 38034911005 en
9
Key
Display presentation Process description
Display of measured value for calibration point 1.
2 calibration points (0 and 1) were recorded. The cali-
bration can be saved with or further calibration
points can be recorded.
Second calibration point at 15 kg weighing range.
Example: 6 kg
Display of measured value for calibration point 2.
Third calibration point at 15 kg weighing range.
Example: 9 kg
Display of measured value for calibration point 3.
Fourth calibration point at 15 kg weighing range.
Example: 12 kg
Display of measured value for calibration point 4.
Fifth calibration point at 15 kg weighing range.
Example: 15 kg
Display of measured value for calibration point 5.
Sixth calibration point at 15 kg weighing range.
Example: 15 kg
Display of measured value for calibration point 6.
Seventh calibration point at 15 kg weighing range.
Example: 15 kg
Display of measured value for calibration point 7.
After acknowledgment of calibration point 7 the device restarts. The scale calibration is
completed.