to take mitigation measures, such as reorienting or relocating Kick 2 or shielding the
location.
Portable and mobile radio frequency communication may affect the equipment.
Other OR equipment may cause an interference with Kick 2, even if it is CISPR emission
compliant.
Restrictions to Environment
Warning
System components are not suitable for use in the presence of flammable anesthetic
mixtures containing air, oxygen or nitrous oxide.
Do not place system components on unstable ground where the system could tip over and be
seriously damaged.
Protection Covers
Keep protection covers in a clean and dry place during surgery.
Related Links
4.4 Parking and Storage on page 58
7.2 Environmental Requirements on page 80
USING KICK 2
System and Technical User Guide Rev. 1.3 Kick 2 Navigation Station 31