3 USING KICK 2
3.1 System Setup
3.1.1 Overview
Installation
The Kick 2 is delivered pre-installed and ready for use.
All system components are suitable for continuous use during surgical procedures.
Interference
Kick 2 does not have an essential performance as defined by IEC 60601-1. However, the
influence of electromagnetic disturbances might lower performance.
Strong electromagnetic interferences may affect the touch technology and the loss of wireless
network connection. If the touchscreen is not reacting properly or creating phantom touch
activations, relocate Kick 2.
The frequency or frequency band of transmission: 2.4 GHz and 5 GHz with a bandwidth of 20
MHz in 802.11 a/b/g standard, 40 MHz in 802.11n and up to 160 MHz in 802.11 ac
Warning
The system produces electromagnetic fields that may interfere with other sensitive
equipment, and can itself be disturbed by other electromagnetic fields.
MR Safety
Warning
Kick 2 has not been tested in an MR environment.
System Position
Ensure that the system is set up in a way that the mains power plugs are easily accessible. In
case of malfunction, you must be able to easily unplug the mains power cables.
Ensure that the system is set up so it is not possible for the patient to touch or come in contact
with the equipment.
Electromagnetic Compatibility
Special precautions regarding electromagnetic compatibility (EMC) must be installed and put into
service according to the EMC information provided in this guide.
USING KICK 2
System and Technical User Guide Rev. 1.3 Kick 2 Navigation Station 27