7.4 Compliances
7.4.1 Electromagnetic Emissions
Electromagnetic Environment
The Kick 2 is intended for use in the electromagnetic environment specified in the table below.
The user is responsible for ensuring that the systems are used in such an environment.
Portable RF Communications Equipment
Warning
Portable RF communications equipment (including peripherals like antenna cables and
external antennas) should be used no closer than 30 cm from any part of the Monitor Cart
including cables specified by Brainlab. Otherwise, degradation of the performance could
result.
Declaration
The Kick 2 is intended for use in professional healthcare facilities only. The following lists all
applicable emission tests and standards, as well as the used compliance levels:
Emissions Test Standard Compliance Level
Conducted and radiated
RF emissions
CISPR11
Group 1
Class A
Harmonic distortion IEC 61000-3-2 Class A
Voltage fluctuations flick-
er
IEC 61000-3-3 Complies
NOTE: The emissions characteristics of this equipment make it suitable for use in industrial areas
and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11
class B is normally required) this equipment might not offer adequate protection to radio-
frequency communication services. The user might need to take mitigation measures, such as
relocating or reorienting the equipment.
Warning
The system should not be used adjacent to, or in direct contact with other equipment. If
this cannot be avoided, normal operation must be verified in the configuration in which it
will be used.
COMPLIANCES AND SPECIFICATIONS
System and Technical User Guide Rev. 1.3 Kick 2 Navigation Station 83