6.4.2 Testing Medical Electrical Systems
Measuring Devices
The measurement instructions provided are an example for one measuring device.
The individual steps may vary from one measuring device to another. Refer to the corresponding
manufacturer's instructions for the measuring device you are using.
Testing Precautions
The described test only covers the Brainlab equipment and not the connected equipment.
Please additionally follow the description given by the manufacturer of the connected device.
All devices shall be in operation mode: During the measurement all mains switches shall be in
operating position (ON).
Tests shall be performed in normal condition as well as in single fault condition.
Test Guidelines
When performing the test, keep the following points in mind:
Perform required test steps in the order defined above
Measure all areas of the system indicated in the following chapters
All tests must be passed
The calibration of the measuring device must be valid at the time of measurement
Perform all tests according to IEC 60601-1:2005
NOTE: The measurement outlet connection of the measuring device must remain established
during the entire measurement cycle. If you unplug the measurement outlet during measurement,
the measuring device recognizes that the connection is no longer established and you will need to
repeat the entire test.
If a test does not pass, Kick 2 must be repaired by Brainlab support. After repair, repeat the entire
electrical safety test from the beginning.
Safety Inspection Form
Print out or make a copy of the safety inspection form.
Enter the inspection results.
Keep it as a record of the inspection.
Required Test Steps
Step
1. Perform visual inspection.
2. Check touch current.
3. Perform functional test.
4. Report and evaluate results.
5. Check and prepare system for normal use.