6.2.2 Test Steps
Measuring Devices
The measurement instructions provided are an example for one measuring device.
The individual steps may vary from one measuring device to another. Refer to the corresponding
manufacturer’s instructions for the measuring device you are using.
NOTE: If you are unsure whether your measuring device is suitable, contact Brainlab support.
Test Guidelines
When performing recurrent tests:
• Perform required test steps in the order defined above.
• Measure all areas of the system indicated in the following chapters.
• All tests must be passed before the device can be considered safe.
• The calibration of the measuring device must be valid at the time of measurement.
• Perform all tests according to IEC 62353.
NOTE: The measurement outlet connection of the measuring device must remain established
during the entire measurement cycle. If you unplug the measurement outlet during measurement,
the measuring device recognizes that the connection is no longer established and you will need to
repeat the entire test.
If a test does not pass, device must be repaired by Brainlab support. After repair, repeat the entire
electrical safety test from the beginning.
Safety Inspection Form
• Print out or make a copy of the safety inspection form.
• Enter the inspection results.
• Compare measured values with critical values to determine if the test has been passed.
• Keep form as a record of the inspection.
Related Links
6.2.3 Safety Inspection Form – Recurrent Tests on page 70
Required Test Steps
Step
1. Perform visual inspection.
2. Check protective earth resistance.
3. Check equipment leakage current.
4. Perform functional test.
5. Report and evaluate results.
6. Check and prepare for normal use.
ELECTRICAL SAFETY
System and Technical User Guide Rev. 1.3 Kick 2 Navigation Station 69