3.5 Sterile Use
3.5.1 Sterile Kick Monitor Drape
General Information
Kick 2 is unsterile. The Monitor Cart may be used within the patient environment when using the
Kick Monitor Drape provided by Brainlab.
The drape allows you to intraoperatively access all software functions and position monitor during
surgery without compromising the sterile field.
Do not allow any system parts to enter the sterile field.
Packaging
The Kick Monitor Drape is delivered sterile. Ensure that drape packaging does not have any
holes or tears.
Before use, check the expiration date on drape packaging. If the date has expired, dispose of the
drape and do not use it.
The drape is delivered sterile. If the outer side of the drape comes in contact with an unsterile
environment during unpacking or clinical use, it must be disposed of.
Do not use drape if packaging is broken or torn.
Draped Kick
Figure 8
Operating Draped Touchscreen
Do not touch the drape with sharp tools.
To keep the field sterile, only touch the draped parts of the touchscreen and monitor housing.
Disposal
Dispose of the drape after use.
USING KICK 2
System and Technical User Guide Rev. 1.3 Kick 2 Navigation Station 35