4.4.4 Long Term Storage
General Information
Before and after placing the system into storage for an extended period of time, steps must be
taken to ensure proper functioning.
Disposal
If you do not intend to use the system again, refer to the disposal information.
Related Links
1.2 Legal Information on page 8
Storage for 1-6 Months
Before storing system for one to six months:
Step
1. Unplug all external cabling and store it together with the equipment.
2. Cover monitor with protection cover.
3. Store equipment in a dry place with a temperature between 21°C (70°F) to 30°C (86°F).
Before using equipment that has been stored for approximately three months:
• Clean the equipment.
• Start system and verify correct function before using it for surgery.
Related Links
5.1 Cleaning the Monitor Cart on page 63
Storage for 6 Months or Longer
Before storing system for six months or longer:
Step
1. Unplug all external cabling and store it together with the equipment.
2. Cover monitor with protection cover.
3. Store equipment in a dry place with a temperature between 21°C (70°F) to 30°C (86°F).
Before using equipment that has been stored for six months or longer:
• Clean the equipment.
• Perform a recurrent test to verify safety and effectiveness of equipment (offered by Brainlab).
• Start system and verify correct function before using it for surgery.
Related Links
6.2 Recurrent Test Requirements on page 68