6.3.2 Equipment Leakage Current
General Information
Connect the measurement tip of your measuring device with the conductive parts shown in the
table below and measure the leakage current.
Equipment Leakage Current Measurement Points
Point Tested Measured Value Test
Passed?
Potential equalization port
Microscope fixation screw
Screw on the back of the
monitor (any screw under
flap)
Post fixation
ELECTRICAL SAFETY
System and Technical User Guide Rev. 1.3 Kick 2 Navigation Station 73