EasyManua.ls Logo

Brookfield DV3T - A.4 Verifying Calibration

Brookfield DV3T
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Brookeld Engineering Labs., Inc. Page 93 Manual No. M13-167-A0415
A.4 Verifying Calibration
1. Determinetheappropriatesamplevolume.RefertoTableA-1todeterminethecorrectsample
volumerequiredforthespindletobeutilized.
2. SelectaBrookeldViscosityStandarduidthatwillgiveviscosityreadingsbetween10%and
100%offullscalerange.RefertoAppendix Bforviscosityrangesofconespindles.
BrookeldusesmineraloilviscositystandarduidstocalibrateWellsBrookeldCone/Plate
Rheometersatthefactory. Brookeld recommends that customers use mineraloilviscosity
standarduidswhenyouperformacalibrationcheck.
Ifyoudecidetouseasiliconeviscositystandarduid,donotuseauidwithaviscosityvalue
greaterthan5000cPwithaCone/Plate.Brookeldoffersacompleterangeofmineraloilviscosity
standardssuitableforusewithCone/Platesforviscositiesabove5,000cPorshearratesabove
500 sec
-1
;seeTableE-2inAppendixFforalistofavailableuids.
Itisbesttouseaviscositystandarduidthatwillbeclosetothemaximumviscosityforagiven
conespindle/speedcombination.
Example: DV3TLVRheometer,ConeSpindleCPA-42Z,BrookeldSiliconeViscosityStandard
havingaviscosityof9.7cPat25°C
At60RPM,thefullscaleviscosityrangeis10.0cP.Thus,theRheometerreadingshouldbe
97%torqueand9.7cPviscosity±0.197cP.Theallowableerror(±0.197cP)isacombination
ofRheometeraccuracyanduidtolerance(refertoInterpretation of Calibration Test Results
inAppendixF).
3. Withthemotoroff,removethesamplecupandplacetheviscositystandarduidintothecup.
Table A-1
Cone Part No.
CPA-40Z, CPE-40, CP-40
CPA-41Z, CPE-41, CP-41
CPA-42Z, CPE-42, CP-42
CPA-51Z, CPE-51, CP-51
CPA-52Z, CPE-52, CP-52
Sample Volume
0.5 mL
2.0 mL
1.0 mL
0.5 mL
0.5 mL
4. AttachthesamplecuptotheRheometerandallowsufcienttimeforthesample,cupandcone
toreachtemperatureequilibrium.
5. Turnthemotoron.Setthedesiredspeed(s).Measuretheviscosityandrecordthereadingin
both%torqueandcentipoise(cP).
6. Verifythattheviscosityreadingiswithintheallowable1%deviation,asexplainedearlier,for
thespecicviscositystandarduid(s)thatyouareusing.
 TheCPEorCPAdesignationontheconespindleindicatesusewithElectronicGapSettingCone/
PlateRheometers/Rheometersonly.

Table of Contents

Related product manuals