Maintenance and troubleshooting
10.3. Replacing the probe
DANGER
Risk of injury due to high pressure in the installation.
Stop the circulation of fluid, cut off the pressure and drain the pipe before loosening the process connections.
Risk of injury due to electrical voltage.
Shut down and isolate the electrical power source before carrying out work on the system.
Observe all applicable accident protection and safety regulations for electrical equipment.
Risk of injury due to high fluid temperatures.
Use safety gloves to handle the device.
Stop the circulation of fluid and drain the pipes before loosening the process connections.
Risk of injury due to the nature of the fluid.
Respect the regulations on accident prevention and safety relating to the use of aggressive fluids.
NOTE
The transmitter loses the IP67 and IP65 protection ratings when dismounted.
Protect the inside of the dismounted transmitter.
Remove the probe as shown below.
Unscrew the nut between the sensor holder and the electronic
module.
Remove the electronic module by pulling it straight out. There
may be a slight resistance due to the seal.
English
Type 8202 ELEMENT

Table of Contents

Other manuals for Burkert 8202 ELEMENT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Burkert 8202 ELEMENT and is the answer not in the manual?

Burkert 8202 ELEMENT Specifications

General IconGeneral
BrandBurkert
Model8202 ELEMENT
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals